Reproducibility and Accuracy of 2D High Resolution Myocardial TRAFF2
Back to course
[Please upgrade your browser to play this video content]
Video Transcription
Video Summary
The study evaluated myocardial T-RAF2 mapping, a low-SAR rotating-frame MRI method, on a clinical 3T scanner. Using phantom and healthy-volunteer tests, the team found the sequence feasible, accurate, and highly reproducible. Maps showed good sensitivity to tissue composition, homogeneous myocardium values, and no visible B1+ or B0 artifacts. A 3-parameter fit model generally gave lower, more consistent values than a 2-parameter fit. Overall, T-RAF2 appears promising for non-contrast cardiac scar assessment with clinically acceptable SAR.
Keywords
myocardial T-RAF2 mapping
rotating-frame MRI
clinical 3T scanner
cardiac scar assessment
low-SAR MRI
×