Video Transcription
Video Summary
The speaker explains how MRI measures blood flow using phase-contrast imaging. Spin phase shifts caused by motion are converted into velocity, and flow is calculated from vessel area and velocity over the cardiac cycle. Key topics include choosing the right VENC to avoid aliasing, correcting background phase errors, shimming, and placing vessels in isocenter. Accurate volume flow requires slow, non-segmented sequences, while breath-hold scans are acceptable for peak velocity. The talk also emphasizes scanner-specific validation using healthy volunteers and comparing aortic and pulmonary flows for consistency.
Keywords
MRI blood flow
phase-contrast imaging
VENC aliasing
volume flow measurement
background phase correction
×